Miniature Head Mount Temperature Transmitters with RFID Communications

TX400-Series

Head Mount Temperature Transmitters | TX400-Series

Head Mount Temperature Transmitters

86,840원 TX401

2 year warranty CE
  • UL Approved
  • High Accuracy & 16-Bit Conversion
  • TX401 Accepts Pt100, Pt1000 or Ni100 RTD Inputs
  • TX402 Accepts Pt100 or Ni100 RTD Inputs and Types J/K/T/E/R/S/B/N Thermocouples
  • User Configurable via TX400-RFID (NFC) Programmer with PC Software or Free Download Google App - "Direct Link" By Phone
  • Transmitters Have Non-Volatile Memory (Circular Buffer) for Data-Logging with User-Selectable Sampling Time
프로세스 측정 / 제어 장비 - 관련 제품 보기

설명

The TX401 and TX402 head mount temperature transmitters transform a temperature signal acquired from a Pt100, Pt1000 (TX401 only) or Ni100 RTD sensor or from a thermocouple (TX402 only) into a linearized 2-wire loop-powered 4 to 20 mA output. The characteristics of these converters ensure high precision on the reading scale with 16-bit conversion. The 4 to 20 mA output can be scaled based on the desired temperature input range.

The programming procedure uses an RFID (NFC) mode with the dedicated TX400-RFID programmer that allows the user to make all calibrations and settings quickly and without the need to power and connect up the transmitter. Simply connect the TX400-RFID programmer to the USB port of your PC, start the RF Programmer configuration software and place the transmitter on top of the TX400-RFID programmer to establish communications.

An additional programming tool is the Omega Brand Google App - "Direct Link" relying on RFID/NFC connectivity and allowing straightforward programming without wirings by Android devices.

These temperature transmitters are also provided with a data logging function for the input signal. Transmitters have a non-volatile memory (circular buffer) for data-logging with user-selectable sampling time. The user selects the sampling time (1 to 3600 seconds) and then each time the 4 to 20 mA loop powers the device, the input value is stored to nonvolatile memory.

Using the TX400-RFID programmer along with the RF Programmer configuration software (software is a free download from OMEGA) the user can:
     Completely configure the transmitter including scaling the 4 to 20 mA output to desired temperature input range
     Implement field calibration to compensate for gain and offset errors
     Download logged data to PC for visualization and printing

Specifications
RTD Input Types and Ranges:
    TX401, TX402: Pt100 (a = 0.00385), 2, 3, or 4-wire connection: TX401: -200 to 800°C (-328 to 1472°F), TX402: -200 to 600°C (-328 to 1112°F)
    TX401 Only: Pt1000 (a = 0.00385), 2-wire connection; -200 to 800°C (-328 to 1472°F)
    TX401, TX402: Ni100 (a = 0.00618), 2, 3, or 4-wire connection: TX401: -50 to 170°C (-58 to 338°F), TX402: -60 to 180°C (-76 to 356°F)
Thermocouple Input Types (TX402 Only): J/K/T/E/R/S/B/N/mV
Output Resolution: 1 µA
Over-Range Output: FS + 5°C
Under-Range Output: FS - 5°C
Output Error (Failure): Selectable: TX401: between 21.5 or 3.8 mA, TX402: between 21 or 3.8 mA
Current Output Protection: 30 mA approx
Rejection: 50 to 60 Hz
Maximum Transmission Error: 0.1% FS or 0.2°C whichever is greater
Sampling Time: 300 msec
Response Time (10 to 90%): 600 msec
Non-Volatile Memory (Circular Buffer) for Data Logging:
    TX401: 3966 Data Points
    TX402: 2546 Data Points
Cable Resistance: 20Ω max
Temperature Coefficient: <100 ppm
Power: 2-wire loop powered; operating range 6 to 32 Vdc
Isolation:
    TX401: Non-isolated
    TX402: galvanic, 1 KVac input/output
Operating Environment: -40 to 85°C (-40 to 185°F), 30 to 90% RH (non condensing)
Storage Temperature: -40 to 85°C (-40 to 185°F)
Mounting: DIN/B head
Enclosure Material: Nylon (PA66)
Enclosure Rating: NEMA 1 (IP20) according to CE, EN 61000-6-4, EN 61000-6-2
Programming: Wireless with RFID technology (NFC)
Software: Windows® XP/7/8 (32-bit and 64-bit), Windows 10 (64-bit)
Connection: Screw terminals
Weight: 30 g (1.05 oz) approx
Dimensions: 23 H x 45 mm dia (0.9 x 1.77")
가격 확인(원하는 수량을 기입하십시오.)

다른 고객님이 함께 구매하신 제품 : 이 아이콘을 클릭하시면 기존 고객들께서 이 제품을 주문할 때 함께 구매하신 제품리스트가 보입니다.

제품 번호/Desc.
수량
설명
Click to expand/minimize a list of what others bought with theTX401
TX401
86,840원
배송: 2 주
 
RTD input head mount RFID temperature transmitter
Click to expand/minimize a list of what others bought with theTX402
TX402
103,540원
배송: 2 주
 
Thermocouple/RTD input head mount RFID temperature transmitter
Click to expand/minimize a list of what others bought with theTX400-RFID
TX400-RFID
245,930원
배송: 2 주
 
RFID programmer
모든 가격은 원화로 표시되어 있습니다.
참고:
TX400-RFID programmer is supplied with complete user’s guide and USB interface cable. User’s guide for the TX401 and TX402 transmitters can be downloaded from OMEGA. RF programmer PC configuration software is a free download from OMEGA.

Google App "Direct Link" can be free downloaded from Google Store or get a copy from our software link here.


주문 예: (1) TX400-RFID RFID programmer, 245,930원  plus (1) TX402 Thermocouple//RTD input head mount RFID temperature transmitter, 103,540원, 245,930 + 103,540 = 349,470원

제품 메뉴얼:

다운로드 Head Mount RTD Input Transmitter RFID
다운로드 Head Mount TC/RTD Input Transmitter RFID
다운로드 RF Programmer for TX400, TXDIN400 Series

소프트웨어:

다운로드 RF Programmer Software for TX400 and TXDIN400
다운로드 Omega Brand NFC Apps – Direct Link
We noticed you are from the US. Visit your local site for regional offers and live support.